APT3100 Smart Pressure Transmitter Operation Manual M3100-EO1G
21
Duon System Co.,Ltd.
- Change Upper Range Value
- Executing the menu to put Zero+Span button.
- Moving to the next menu to put Zero button when 1 TRIM message appear.
- Moving to the sub directory to put Span button when 2 Setup message appear.
- Moving to the next menu to put Zero button when 21 Unit message appear.
- Executing the function to put Span button when 22 U-RNG message appear.
- The procedure of setting value is same as Zero Adjustment.
- Change Lower Range Value
- Executing the menu to put Zero+Span button.
- Moving to the next menu to put Zero button when 1 TRIM message appear.
- Moving to the sub directory to put Span button when 2 Setup message appear.
- Moving to the next menu to put Zero button when 21 Unit message appear.
- Moving to the next menu to put Zero button when 22 Unit message appear.
- Executing the function to put Span button when 23 L-RNG message appear.
- The procedure of setting value is same as Zero Adjustment.
- Decimal Place
- Executing the menu to put Zero+Span button.
- Moving to the next menu to put Zero button when 1 TRIM message appear.
- Moving to the next menu to put Zero button when 2 Setup message appear.
- Moving to the sub directory to put Span button when 3 LCD message appear.
- Executing the function to put Span button when 31 DEC-PL message appear.